Contact Us | Customer Service | (800) 635-9729 |
   

Radiology Supplies

Digital kVp Meter

SKU# GP-KVP-METER

Be the first to review this product

Please log on or call 800-635-9729 for best pricing and availability.

Product Name
SKU: 122499 - MFR: 07-494 - Wide-Range Digital kVp Meter
SKU: 112003 - MFR: 07-492 - Mammographic Digital kVp Meter

 

Digital kVp Meter

Details

Digital kVp Meter

Digital kVp Meters:  Wide-Range or Mammographic

Whether you choose the Wide-Range, Mammographic, or the Dental Digital kVp Meter, you will get quick and accurate measurements of your diagnostic x-ray generator tube potential. These instruments need no connection to the x-ray generator.

These lightweight, rugged units are extremely easy to use: simply place on the x-ray table, with the detector facing the x-ray source. With the beam’s central ray centered on the detector, an exposure is made, and the reading appears immediately on the large, easy-to-read liquid crystal display. Unique features are provided to ensure maximum efficiency and accuracy. Readings remain on display until the next exposure is made, at which time the reading is automatically updated. Automatic display indicators tell you when adjustment of exposure factors or battery replacement is necessary. Neither remote-control cables nor time-consuming manual re-zeroing are needed. A BNC connector is provided for radiation waveform display on a storage oscilloscope.

Features

  • Automatic display reset
  • No remote control cables
  • Scope output for waveform analysis
  • Compact, lightweight and battery operated
  • Choose from three kVp meters
  • Easy read digital display
  • Painless setup

Product Specifications
Ranges
Wide Range: Low: 50 kVp to 90 kVp, 0.1 kVp resolution
High: 80 kVp to 150 kVp, 0.1 kVp resolution
Mammographic: 24 kVp to 40 kVp, 0.1 kVp resolution
Dental: 45 kVp to 90 kVp, 0.1 kVp resolution
Accuracy
Wide Range: ± 3% or 3 kVp (whichever is greater)
Mamographic: ± 3% or 1.5 kVp (whichever is greater)
Dental: ± 3% or 3 kVp (whichever is greater)
mAs requirements
Wide Range: (45.7 cm SDD) 18 mAs at 120 kVp; 50 mAs at 60 kVp, single phase. Minimum exposure time 1/20 (0.05) sec
Mammographic: (25 cm SDD) 100 mAs at 24 kVp. Minimum exposure time 1/20 (0.05) sec
Dental: 8.5 mAs at 45 kVp; 0.026 mAs at 90 kVp
Controls
Wide Range: On/off, single/three-phase and range selection switch
Mammographic: On/off and Moly/Tungsten selector switches
Dental: On/off and single/three-phase selector switches
Operating temperature

10 ºC to 40 ºC (50 ºF to 104 ºF)
Relative humidity

0% to 90%, noncondensing
Power requirements

9 V alkaline battery, 150 hours operation
Display

3.5 x 0.5 inch LCD. Automatic indication of (a) low battery condition, (b) need to adjust exposure factors
Output signal

BNC connector for waveform analysis
Dimensions(WxDxH)

20 cm x 15 cm x 6 cm (8 in x 6 in x 2.5 in)
Weight

0.9 kg (2 lb)

Additional Information
SKU GP-KVP-METER
Manufacturer FLUKE ELECTRONICS CORPORATION
Packaging EACH
BAAN SKU No
Short Description Digital kVp Meter