Mammography Phototimer Consistency Test Tool

Details

Mammography Phototimer Consistency Test Tool

set of four acrylic slabs (10 x 12.5 x 2 cm thick). Weight: 3 lb (1.34 kg)

Introduction
The mammographic unit’s automatic exposure control should be capable of maintaining optical density within ± 0.15 OD as the voltage is varied from 25 to 35 kVp, and as breast thickness is varied from 2 to 8 cm for each technique. Test images taken of uniform phantoms of varying thicknesses should not differ by more than 0.30 OD from each other. These tests should be carried out over the kVp range customarily used by the mammography center.

The Phototimer Consistency Test Tool is available in two materials: acrylic; and, for more accurate results, breast-tissue equivalent BR-12 material. Both are supplied in uniform 2.0 cm slabs to produce thicknesses of 2, 4, 6, and 8 cm.

Applications
Available in either acrylic or tissue-equivalent BR-12 material*
Should be used to test thickness tracking
The American College of Radiology recommends this type of product in their quality assurance program
Meets MQSA requirements

Additional Information